Spectral Calibration

Wavelength calibration of a Czerny–Turner spectrograph from mercury lines: linear, quadratic and cubic fits against the grating-equation pixel map.

Model: Textbook approximation

Detector Range (grating equation)

394.0 – 605.2 nm

α = 33.09°, β₀ = 3.09°

Dispersion at Centre (theory)

0.1040 nm/px

d·cos β₀/(m·f) × pixel pitch

Linear Fit λ = a·i + b

a = 0.103557 nm/px

b = 393.8798 nm at pixel 0

Lines on Detector

6

degree k needs k + 1 lines

Fit Quality

DegreeRMS residual at lines (nm)Max |fit − true| over detector (nm)Dispersion at centre (nm/px)
10.09830.6250.103557
20.09040.4940.103526
30.003190.01200.104005

Calibration Error λ_fit − λ_true

Error of each fitted polynomial against the grating-equation truth at every pixel; dots mark the calibration lines. Outside the outermost lines the fit extrapolates.

Model

x = (i − (N − 1)/2)·p, β = β₀ + atan(x/f), λ = d(sin α + sin β)/m

λ_fit(i) = Σ c_k·u^k, u = (i − (N − 1)/2)/((N − 1)/2), least squares over the lines

The true pixel-to-wavelength map is the grating equation for a Czerny–Turner spectrograph with equal focal lengths and a flat detector normal to the central ray, set by the centre wavelength and the fixed deviation α − β₀ (as on the spectral range page). Both sin β and the atan of the flat detector make it non-linear. When β₀ is small the distortion is nearly odd about the centre, so a quadratic barely improves on a line and a cubic is needed. Each line centroid is offset by Gaussian noise of the given σ (seeded, so the page is reproducible). Aberrations, detector tilt and focal-plane curvature are not modelled. Enter air wavelengths: the grating equation holds for the wavelength in the air around the grating.

References: C. Palmer, E. Loewen, Diffraction Grating Handbook, 6th ed. (Newport, 2005), ch. 2; A. Kramida, Yu. Ralchenko, J. Reader and NIST ASD Team, NIST Atomic Spectra Database (Hg I lines); C. J. Sansonetti, M. L. Salit, J. Reader, Appl. Opt. 35, 74 (1996).

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

The true pixel-to-wavelength map is the grating equation on a flat detector normal to the central ray (no aberrations, tilt or focal-plane curvature); line centroids carry seeded Gaussian noise.

  1. Palmer C., Loewen E. (2005). Diffraction Grating Handbook, 6th ed., ch. 2. Newport.
  2. Kramida A., Ralchenko Yu., Reader J., NIST ASD Team. NIST Atomic Spectra Database (Hg I air wavelengths). doi:10.18434/T4W30F
  3. Sansonetti C. J., Salit M. L., Reader J. (1996). Wavelengths of spectral lines in mercury pencil lamps. Appl. Opt. 35, 74–77. Source of the ASD Hg I lines, ±0.1 pm. doi:10.1364/AO.35.000074
Report a problem

A wrong value, a formula, a reference, or a page that misbehaves. The report includes this page and its current inputs.