Multiple Wavelength MPE

Eye exposure to a beam with several wavelengths on the ICNIRP 2013 limits: lines absorbed in the same tissue add, while the retina and the cornea count independently.

Model: Exact

Educational only — not compliance-grade

These pages implement the published exposure limits for the cases each one states, but they are educational tools: no laser safety officer has signed them off.

  • Do not use these results alone for PPE selection, NHZ / NOHD approval, controlled-area sign-off, enclosure/interlock approval, or formal classification.
  • Always verify against ANSI Z136.1, IEC 60825-1, your local procedures, and the responsible laser safety officer / CLSO.

0.25 s (the blink reflex) only protects against visible beams; use 10 s or more when a line is invisible.

Retina: sum of shares

1.22

Exceeds the limits

Anterior eye (cornea and lens): sum of shares

0.637

Within the limits

Whole beam

Exceeds the limits

Attenuate every line by 1.22× (OD 0.09) at least

Retina

Anterior eye (cornea and lens)

450 nm520 nm638 nm1550 nmwhite dashed line: the limit (sum = 1)
λ (nm)PowerLimitLargest power aloneShare (P / largest)
450400 µWRetina, thermal980 µW0.408
Retina, photochemical (blue light)applies from 10 s0
520400 µWRetina, thermal980 µW0.408
Retina, photochemical (blue light)applies from 10 s0
638400 µWRetina, thermal980 µW0.408
155010.0 mWCornea (IR)15.7 mW0.637
Same limit, all linesRetina, thermal: 1.22 · Retina, photochemical (blue light): 0 · Cornea (IR): 0.637

ICNIRP 2013 (Health Phys. 105:271, p. 279): wavelengths absorbed in the same tissue add; wavelengths absorbed in different tissues, such as the retina and the cornea, are assessed independently. Each line's share of a limit is its power over the largest power that stays within that limit for every exposure up to t, averaged over the limit's aperture (7 mm for the retina; 1–3.5 mm for the cornea; 3.5 mm for the anterior-segment limit at 1150–1400 nm). A beam under 1 mm (1/e diameter, d/√2 for the 1/e² d) is held to its peak on the cornea below 0.35 s (ICNIRP 2013 Table 5 note c). A line's share of a tissue is its largest share there (one line's thermal and blue-light retinal limits are met separately); the shares then add over the lines, and each tissue's sum must stay at or below 1.

All lines are taken as one co-aligned beam of the diameter above, viewed for the same time. The retinal limits are for a point source (the conservative case). Not modelled: skin, pulses and pulse trains.

Model and references

Exact. The equations are exact within the stated assumptions, or are a standard's tabulated values. The results are as good as the inputs. About the tiers

ICNIRP 2013 eye limits per line (point source, averaged over each aperture or, under 1 mm (1/e) on the cornea below 0.35 s, the peak (Table 5 note c), exposures up to t); shares add per tissue (retina; cornea and lens) as on p. 279. Not modelled: skin, pulses, lines of different beam size.

  1. ICNIRP (2013). Guidelines on limits of exposure to laser radiation of wavelengths between 180 nm and 1,000 µm. Health Phys. 105(3), 271–295. p. 279 (multiple wavelengths), Tables 5 and 8. doi:10.1097/HP.0b013e3182983fd4