SPAD Dead Time

Dead time effects on measured count rates, pile-up loss, and correction for SPAD detectors.

Model: Textbook approximation

Results

Measured Rate (nonparalyzable)
8.000 Mcps
Count Loss = dead fraction of the time
20.00%
Correction Factor R/m
1.2500
True Rate at 10 % Loss
4.44 Mcps

Pile-up Correction Table (τ = 25 ns)

True rate (Mcps)Measured NPMeasured PLoss NPLoss P
10.9760.9752.4%2.5%
54.4444.41211.1%11.8%
108.0007.78820.0%22.1%
2013.33312.13133.3%39.3%
5022.22214.32555.6%71.3%
10028.5718.20871.4%91.8%

Measured vs True Rate

Count Loss vs Dead Time

Formulas

Non-paralyzable: m = R / (1 + R·τ); correction R = m / (1 − m·τ)

Paralyzable: m = R·exp(−R·τ); peak m = 1/(e·τ) at R = 1/τ

Loss 1 − m/R equals the fraction of time the counter is dead

Model and references

Textbook approximation. A standard approximation from the literature (for example paraxial, plane-wave or Gaussian-beam), valid within the range the page states. About the tiers

The two limiting dead-time models for Poisson arrivals. Real SPADs with active quenching sit near the non-paralyzable case, passively quenched ones between the two, and the dead time itself may vary.

  1. Knoll G. F. (2010). Radiation Detection and Measurement, 4th ed. Wiley. Ch. 4, section VII: non-paralyzable m = n/(1 + nτ) and paralyzable m = n·e^(−nτ) dead-time models.
  2. Müller J. W. (1973). Dead-time problems. Nucl. Instrum. Methods 112, 47–57. Dead-time models and the statistics of the recorded counts. doi:10.1016/0029-554X(73)90773-8